Setup Load Cell:
| Parameter | Current | Updated |
|---|---|---|
| MEASURE TYPE | - | |
| CALIBRATION MODE | - | |
| MEASURE UNIT | - | |
| CELL SENSIBILITY [mV/V] | - | |
| CELL FULLSCALE [g] | - | |
| REFERENCE WEIGHT [g] | - | |
| FILTER LEVEL | - | |
| DENOISE FILTER | - | |
| RESOLUTION MODE | - | |
| STABLE DELTA WEIGHT [g] | - | |
| STABLE DELTA TIME [x100ms] | - |
β Stable Weight Detection (Recommended):
Set STABLE DELTA WEIGHT = 2.0g and STABLE DELTA TIME = 10 (1 second) to eliminate weight jumps. Only stable readings will be sent to the monitor.
Tip: For 1g steps, also set RESOLUTION MODE = MANUAL RESOLUTION with value 1.0
π For Maximum Responsiveness (No Filter):
Set FILTER LEVEL = 0 (2ms) and DENOISE FILTER = 0 (Disabled) for the fastest, most responsive weight readings with no filtering applied.
Live Weight Monitor (Real-Time):
OFF
SCALE WEIGHT
Unstable
Connection:
Disconnected
Updates:
0
Rate:
-- Hz
Uptime:
0s
TEST AND LOAD CELL CALIBRATION:
| Measurement | Value | Unit |
|---|---|---|
| NET WEIGHT | - | [g] |
| GROSS WEIGHT | - | [g] |
| TARE WEIGHT | - | [g] |
| ADC FILTERED | - | |
| PIECES COUNT | - |
STATUS REGISTER
STABLE WEIGHT
FULLSCALE CELL
NEGATIVE WEIGHT
LOAD CELL CALIBRATION
Calibration Procedure:
- Step 1: Ensure reference weight is configured in Setup Load Cell tab
- Step 2: Ensure scale platform is empty and stable, then click "Tare Acquisition"
- Step 3: Place the reference weight on the scale and wait for stability
- Step 4: Once stable, click "Accept Reference Weight" to complete calibration